![Rohde & Schwarz 1311.6010.22 User Manual Download Page 17](http://html.mh-extra.com/html/rohde-and-schwarz/1311-6010-22/1311-6010-22_user-manual_1477768017.webp)
Firmware V3.12
R&S
®
ZNB/ZNBT
17
User Manual 1173.9163.02 ─ 53
3
Firmware V3.12
Version 3.12 of the R&S
ZNB/ZNBT firmware provides the following changes:
Solved issues
●
Firmware versions 3.x created incompatible
*.calkit
files
(
*.calkit
files containing snp data could not be used with previous FW versions)
●
R&S
ZNBT8 with CPU board "LPW10" and image version "1.0 (64 Bit)": update to
firmware versions 3.x failed
CPU board and image version can be found in the "Image Information" section of
the
●
PC firmware simulation failed to run due to missing Microsoft Visual C++ redistrib-
utable
Firmware version
► To check your R&S
ZNB/ZNBT firmware version, select "Help" > "About..." from the
main menu.
3.1
Changes in Firmware Version 3.11
Solved issues
●
Problems with fan control
●
Impedance normalization failed for some special balanced port configurations
●
User characterization of R&S
ZN-Z154 did not work
●
Firmware simulation: trace noise could not be disabled in Simulation Config tool
3.2
Changes in Firmware Version 3.10
New functionality
●
Ground loop de-/embedding per port group (i.e. per DUT)
See
●
New "Distance to Fault Measurements" software option R&S
ZNB-K3/R&S
ZNBT-
K3
See
Chapter 5.7.3, "Distance to Fault Measurements"
New remote control features
●
Memory-mapped trace data transfer
See
Chapter 5.4.3, "Memory-Mapped Trace Data Transfer"
Improvements
●
Limit line enhancements:
–
Formula-defined limit lines
See
Changes in Firmware Version 3.10