Command Reference
R&S
®
ZNB/ZNBT
1077
User Manual 1173.9163.02 ─ 53
8.3.14.5
[SENSe:]CORRection:CKIT...
The
[SENSe:]CORRection:CKIT...
commands deal with calibration kits and cal kit
data. The calibration kits are distinguished by their names (<CalkitName>), the optional
labels (<label>) can be used to carry information about the calibration standard.
In order to handle several identical calibration kits with different serial numbers use the
commands of
Chapter 8.3.14.6, "[SENSe:]CORRection:CKIT... with Labels"
[SENSe:]CORRection:CKIT:CATalog?
............................................................................ 1077
[SENSe:]CORRection:CKIT:DELete
............................................................................... 1077
.............................................................................. 1078
................................................................................. 1078
[SENSe:]CORRection:CKIT:SELect
............................................................................... 1079
[SENSe:]CORRection:CKIT:STANdard:CATalog?
............................................................ 1079
[SENSe:]CORRection:CKIT:STANdard:DATA?
................................................................ 1079
[SENSe:]CORRection:CKIT:<ConnType>:SELect
............................................................ 1080
[SENSe:]CORRection:CKIT:<StandardType>
.................................................................. 1081
[SENSe:]CORRection:CKIT:CATalog?
[<ConnectorType>]
Returns a list of all cal kits for a given connector type or for all connector types.
Query parameters:
<ConnectorType>
Name of the connector type (optional). Use
]CORRection:CONNection:CATalog?
names.
If omitted, the command returns the list of all cal kits.
Example:
See
[SENSe<Ch>:]CORRection:CONNection:CATalog?
Usage:
Query only
Manual operation:
See
[SENSe:]CORRection:CKIT:DELete
<CalKitName>
Deletes an imported or user-defined cal kit.
Note:
It is not possible to modify or store predefined or ideal kits.
Setting parameters:
<CalKitName>
String parameter containing an imported or user-defined calibra-
tion kit available on the analyzer.
Example:
See
Usage:
Setting only
Manual operation:
See
"Add / Copy / Delete / Standards..."
SCPI Command Reference