Command Reference
R&S
®
ZNB/ZNBT
940
User Manual 1173.9163.02 ─ 53
Example:
*RST; :CALC:TRAN:VNET:SEND:DEEM:PAR:C2? CSC
Query the default capacitance C2 for the Serial C, shunt C cir-
cuit model. The response is 1E-012 (1 pF).
CALC:TRAN:VNET:SEND:DEEM:PAR:C2 CSC, 2.2E-12
Increase the capacitance to 2.2 pF.
Manual operation:
See
CALCulate<Ch>:TRANsform:VNETworks:SENDed:DEEMbedding<PhyPt>:
PARameters:G<Cmp>
<CircuitModel>, <Conductance>
CALCulate<Ch>:TRANsform:VNETworks:SENDed:DEEMbedding<PhyPt>:
PARameters:G<Cmp>?
<CircuitModel>
Specifies the conductance value G in the "shunt L, shunt C" circuit model for single
ended port deembedding.
In the query form, the <Conductance> parameter must be omitted.
Suffix:
<Ch>
.
Channel number
<PhyPt>
Physical port number
<Cmp>
Number of the conductance component in the circuit model. The
total number of conductances depends on the selected circuit
model.
Parameters:
<Conductance>
Conductance G<Cmp> for the specified circuit model.
Range:
-1kS to 1 kS.
Increment: 1 nS (1E-9 S)
*RST:
0 S
Default unit: Siemens (SI unit symbol: S)
Parameters for setting and query:
<CircuitModel>
SHLC
Circuit model whose conductance G<Cmp> shall be set, see
Example:
*RST; :CALC:TRAN:VNET:SEND:DEEM:PAR:L? SHLC; R?
SHLC; C? SHLC; G? SHLC
Query the default component values for the "shunt L, shunt C"
circuit model.
CALC:TRAN:VNET:SEND:DEEM:PAR:G SHLC, 1
Increase the conductance G to 1 Siemens.
Manual operation:
See
SCPI Command Reference