GUI Reference
R&S
®
ZNB/ZNBT
355
User Manual 1173.9163.02 ─ 53
Remote command:
CALCulate:RIPPle:DISPlay:RESult:ALL[:STATe]
Clear Test
Resets the limit check results.
Remote command:
Global Check
See
TTL1 Pass / TTL2 Pass
See
6.6.2.2
Define Ripple Test Dialog
The "Define Ripple Test" dialog defines the ripple limits for the active trace on a range-
by-range basis. A separate ripple limit can be assigned to each range.
Access:
TRACE – [LINE] > "Ripple Test" > "Def. Ripple Test..."
Defining ripple limits with minimum effort
Choose one of the following methods to create and handle ripple limit ranges effi-
ciently:
●
To configure a limit test with only a few ranges, use "Add" and edit each range in
the table individually.
●
Use the "Align All" button to create non-overlapping, contiguous ranges of equal
width.
●
Save your ripple ranges to a file so you can reuse or modify them in later sessions
("Save Ripple Test..., Recall Ripple Test...").
Background information
Refer to
Lines Softtool