GUI Reference
R&S
®
ZNB/ZNBT
253
User Manual 1173.9163.02 ─ 53
CALCulate<Ch>:PARameter:SDEFine
CALCulate<Ch>:PARameter:SDEFine:SENDed
S<out><in>
Selects one of the four elements of the standard 2-port
quantity for the active trace.
Remote command:
CALCulate<Ch>:PARameter:MEASure
"<Trace_Name>", "S11" | "S12" |
"S21" | "S22"
CALCulate<Ch>:PARameter:SDEFine
"<Trace_Name>", "S11" | "S12" |
"S21" | "S22"
All S-Params
With the active logical ports distributed to m DUTs with n
1
, ..., n
m
ports, respectively, the
analyzer creates n
1
2
+ ... + n
m
2
diagrams and displays the full set of
, one
in each diagram. The diagrams are arranged as an n x n grid.
If "All S-Params" would result in more than 100 diagrams being created, a warning dia-
log is opened that lets you cancel the diagram creation.
The reflection coefficients S
ii
appear in the diagrams on the main diagonal, the trans-
mission coefficients S
ij
(i
≠
j) occupy the other diagrams. By default, reflection coeffi-
cients are displayed in Smith diagrams; transmission coefficients in Cartesian dia-
grams "dB Mag" scale.
See also
.
Remote command:
CALCulate<Ch>:PARameter:SDEFine
CALCulate<Ch>:PARameter:DEFine:SGRoup
S-Param Wizard...
Opens a dialog that guides you through the setup of a standard multi-port S-parameter
measurement in a frequency sweep. See
Chapter 6.2.1.2, "S-Parameter Wizard"
Dialog and opener button are not available for R&S
ZNBT, or if a switch matrix is con-
figured.
Balanced Ports...
Opens a dialog that allows you to enable/disable physical ports and to define logical
ports (balanced or unbalanced) in the active channel. See
Note:
If either multiple DUTs are configured (see
Chapter 6.12.2.3, "Define Parallel
on page 527) or the "Fixture Simulator" is disabled for the rela-
Meas Softtool