6
English
Ranger™ Blood/Fluid Warming System Model 245
NOTICE:
1�
Federal law (USA) restricts this device to sale by or on the order of a licensed
healthcare professional�
1�
To avoid device damage:
•
Do not clean the blood/fluid warming unit with solvents� Damage to the case, label, and
internal components may result�
•
Do not immerse the blood/fluid warming unit in cleaning or sterilizing solutions� The unit
is not liquid proof�
•
Do not insert metallic instruments in the blood/fluid warming unit�
•
Do not use abrasive materials or solutions to clean the heater plates�
•
Do not allow spills to dry inside the unit, as this may make it more difficult to clean
the unit�
•
The Ranger blood/fluid warming unit meets medical electronic interference
requirements� If radio frequency interference with other equipment should occur,
connect the unit to a different power source�
Product Description
The Ranger blood/fluid warming system consists of a Model 245 warming unit and a sterile
warming set�
The Ranger blood/fluid
warming unit
The blood/fluid warming unit is a
compact, lightweight, liquid‑resistant
device with a clamp located on the
side for attachment to an I�V� pole (see
Figure 1)� A carrying handle on the top
of the unit makes it easy to transport�
Located on the front panel you will find:
•
Alphanumeric display that
indicates the heater temperature
during normal operation� In an
over‑temperature condition,
the display alternately flashes
a temperature of 43°C or higher and the word “HI�” An audible alert also sounds� In an
under‑temperature condition, the display alternately flashes a temperature of 33°C or lower
and the word “LO�”
•
Alert indicator light that comes on when either an over‑ or under‑temperature
condition occurs�
Ranger blood/fluid warming set
Blood/fluid warming sets include a cassette, tubing and connectors compatible
with hospital standard I�V� blood/fluid administration sets and a bubble trap�
F
low rate
P
riming volume
P
atient line length
Standard Flow Set
150 mL/min
39‑44 mL
76 cm
152 cm (76 cm patient line
plus 76 cm extension)
High Flow Set
100‑500 mL/min
65‑150 mL
152 cm (76 cm patient line
plus 76 cm extension)
Pediatric/Neonate
100 mL/min
20 mL
46 cm
Refer to the instructions provided with each warming set for information on use�
Slot for cassette
Alphanumeric
display
Bubble trap holder
Power switch
I.V. pole
clamp
Alert indicator light
Figure 1
Figure 2
Warming Set
Summary of Contents for Ranger 245
Page 3: ......
Page 5: ......
Page 21: ......
Page 22: ...Français 17 Unité de réchauffement de sang soluté 3M Ranger Modèle 245 Manuel d utilisation ...
Page 23: ......
Page 40: ...Deutsch 35 Ranger Blut und Infusionswärmesystem Modell 245 Bedienungsanleitung ...
Page 41: ......
Page 59: ......
Page 75: ......
Page 76: ...Español 71 Sistema de calentamiento de sangre fluidos Ranger Modelo 245 Manual del operador ...
Page 77: ......
Page 94: ...Nederlands 89 Ranger bloed vloeistofverwarmings systeem Model 245 Bedieningshandleiding ...
Page 95: ......
Page 112: ...Svenska 107 Ranger blod och vätskevärmesystem Modell 245 Användarhandbok ...
Page 113: ......
Page 130: ...Dansk 125 Ranger blod væskevarmesystem Model 245 Betjeningsvejledning ...
Page 131: ......
Page 148: ...Norsk 143 Ranger varmesystem for blod væske Modell 245 Brukerhåndbok ...
Page 149: ......
Page 166: ...Suomi 161 Ranger veren nesteenlämmitysjärjestelmä Malli 245 Käyttöopas ...
Page 167: ......
Page 184: ...Português 179 Sistema de Aquecimento de Sangue Fluidos Ranger Modelo 245 Manual do Utilizador ...
Page 185: ......
Page 202: ...Ελληνικά 197 Ranger Σύστημα Θέρμανσης Αίματος Υγρών Μοντέλο 245 Εγχειρίδιο χειριστή ...
Page 203: ......
Page 221: ......
Page 222: ...Polski 217 System Ranger do podgrzewania krwi płynów Model 245 Instrukcja obsługi ...
Page 223: ......
Page 241: ......
Page 242: ...Magyar 237 Ranger vér és folyadékmelegítő rendszer 245 ös modell Felhasználói kézikönyv ...
Page 243: ......
Page 260: ...Česky 255 Ohřívací systém pro krev tekutiny Ranger Model 245 Návod k obsluze ...
Page 261: ......
Page 277: ......
Page 278: ...Lietuviškai 273 245 modelio Ranger kraujo skysčių šildymo sistema Naudotojo vadovas ...
Page 279: ......
Page 295: ......
Page 296: ...Русский 291 Система подогрева крови и жидкостей Ranger модель 245 Руководство по эксплуатации ...
Page 297: ......
Page 318: ...Türkçe 313 Ranger Kan Sıvı Isıtma Sistemi Model 245 Kullanım Kılavuzu ...
Page 319: ......
Page 335: ......
Page 336: ...中文 331 Ranger 输血 输液升温系统 245 型 操作指南 ...
Page 337: ......
Page 352: ......