Command Reference
R&S
®
ZNC
560
User Manual 1173.9557.02 ─ 13
Example:
Suppose that a power meter no. 1 is configured and USB-con-
nected to the analyzer.
SENSe:CORRection:COLLect:PMETer:ID 1, 2
Select power meter no. 1, to be connected to Port 2, for the
SMARTerCal.
--> Connect the calibration unit between ports 1 and 2 of the ana-
lyzer,
SENSe:CORRection:COLLect:AUTO:PORTs:TYPE
PFNPort, '', 1, 1, 2, 2
Perform an automatic full 2-port (UOSM) calibration at the ana-
lyzer test ports 1 and 2 using the calibration unit's default calibra-
tion kit file and default port assignments (ports 1 and 2 of the cal
unit).
--> Remove the cal unit and connect the power meter to Port 2 of
the analyzer.
SENSe:CORRection:COLLect:AUTO:POWer 2
Perform an additional receiver power calibration at Port 2. This
completes the
Usage:
Setting only
[SENSe<Ch>:]CORRection:COLLect:AUTO:TYPE
<CalType>[, <Characterization>,
<TestPort1>, <TestPort2>...]
Selects and initiates an automatic calibration at arbitrary analyzer and cal unit ports. This
command also selects the calibration type. A progress monitor for the calibration sweeps
is displayed.
Tip:
If the test setup contains a high attenuation the analyzer may fail to detect the cal
unit ports connected to each of its ports. In this case use the extended command
.
If several calibration units are connected, use
to select a unit for the calibration.
Suffix:
<Ch>
.
Channel number of the calibrated channel
Setting parameters:
<CalType>
FNPort | FOPort | OPTPort | FRTRans | FTRans | RTRans |
UTRans | REFL | RSHort | PFNPort
FNPort
–
full n-port (UOSM) calibration
FOPort
–
full one-port ("Refl OSM") calibration
SFTPort
–
separate full two-port calibrations
OPTPort
–
one path two ports calibration
PFNPort
–
for future extensions
SCPI Command Reference