Command Reference
R&S
®
ZNC
694
User Manual 1173.9557.02 ─ 13
Example:
CORR:COLL:METH REFL1
Select a one-port normalization at port 1 with an open standard as
calibration type.
CORR:COLL:SAVE:DEF
Calculate a dummy system error correction for the normalization
at port 1. The dummy system error correction provides the reflec-
tion tracking error term SCORR3.
INIT:CONT OFF; :INIT; *WAI
Stop the sweep to ensure correct transfer of calibration data.
CORR:DATA? 'SCORR3'
Query the dummy system error correction term. The response is
a 1 (written as
1,0
for the real and imaginary part) for each sweep
point (no attenuation and no phase shift between the analyzer and
the calibration plane).
CORR:DATA 'SCORR3',<ASCII_data>
Replace the dummy system error correction term with your own
correction data, transferred in ASCII format.
INIT:CONT ON
Restart the sweep in continuous mode.
Usage:
Event
[SENSe<Ch>:]CORRection:COLLect:SAVE[:DUMMy]
Calculates the system error correction data from the acquired one or two-port measure-
ment results (
), stores them and
applies them to the calibrated channel
<Ch>
. To avoid incompatibilities, older system
error correction data is deleted unless it has been transferred into a cal pool (
).
This command is the R&S ZVR-compatible equivalent of
. It must be used in combination
with the R&S ZVR-compatible commands for calibration method and standard selection;
see example below.
Suffix:
<Ch>
.
Channel number of the calibrated channel.
Example:
CORR:COLL:METH REFL1
Select a one-port normalization at port 1 as calibration type.
CORR:COLL OPEN1
Measure an open standard connected to port 1 and store the
measurement results of this standard.
CORR:COLL:SAVE
Calculate the system error correction data and apply it to the active
channel.
Usage:
Event