GUI Reference
R&S
®
ZNC
283
User Manual 1173.9557.02 ─ 13
4.4.4.4
Channel Config > Pwr Cal Settings
Provides access to all functions for power meter and power calibration data handling
(transmission coefficients). The settings are identical with the "CHANNEL > CAL > Pwr
Cal Settings" settings; see
chapter 4.4.3.12, "Calibration > Pwr Cal Settings"
4.4.5 Trigger
The "Trigger" menu provides trigger and sweep control settings.
4.4.5.1
Trigger > Trigger
Selects the source of the trigger signal and provides additional trigger settings.
Trigger system of the analyzer
The trigger system is used to synchronize the analyzer's actions with events that can be
provided by an internal or external signal or user-generated ("Manual Trigger"). Triggered
measurements are an alternative to the default mode ("Free Run", "Continuous
Sweep"), where the measurement is continuously repeated without fixed time reference.
Any trigger event may start an entire sweep or a part of it. Moreover, it is possible to
switch off the RF source between consecutive triggered measurement sequences, and
to define a delay between trigger events and the measurement sequences.
Output trigger
The R&S ZNC provides a configurable output trigger signal to synchronize external devi-
ces with the measurement. This signal is available at the rear panel connector EXT TRIG
OUT. Configuration of the output trigger signal is a remote control feature
(
TRIGger:CHANnel<Ch>:AUXiliary...
The trigger settings are also valid for calibration sweeps. This means that, in external
trigger mode, the external trigger signal must be available during the system error cor-
rection, too. To start the calibration sweeps without delay, use the "Free Run" trigger type.
Background information
Refer to
chapter 3.1.4.1, "Partial Measurements and Driving Mode"
Access:
CHANNEL > TRIGGER key or Alt + Shift + R
Channel Settings