Command Reference
R&S
®
ZNC
683
User Manual 1173.9557.02 ─ 13
Example:
CALC:LIM:LOW -10, 0, 0, -10
Define two limit line segments covering the entire sweep range.
Two upper limit line segments with default response values are
created in addition.
CALC:LIM:UPP 0, 5, 5, 0
Change the response values of the upper limit line segments.
CALC:LIM:LOW:STAT ON; :CALC:LIM:UPP:STAT ON; :
CALC:LIM:FAIL?
Switch the limit check on and query the result.
Manual operation:
See
CALCulate<Ch>:PARameter:DEFine
<TraceName>, <Result>[, <TestPortNum>]
Creates a trace and assigns a channel number, a name and a measurement parameter
to it. The trace is not displayed. To display a trace defined via
CALCulate<Ch>:PARameter:DEFine
, a window must be created (
ON
) and the trace must be assigned to this window (
); see example below.
Traces must be selected to become active traces; see
.
Note:
The parameter names in this command differ from R&S ZNC conventions; more-
over the parameter list is not complete. The alternative command
uses a complete parameter list with compatible names.
Suffix:
<Ch>
.
Channel number.
<Ch>
may be used to reference a previously
defined channel. If <Ch> does not exist, it is generated with default
channel settings.
Setting parameters:
<TraceName>
Trace name, e.g. 'Trc4'. See "Rules for trace names" in
<Result>
S11 | S12 | S13 | S14 | S21 | S22 | S23 | S24 | S31 | S32 | S33 |
S34 | S41 | S42 | S43 | S44 | A | B | C | D | R1 | R2 | R3 | R4 | AB |
AC | AD | BA | BC | BD | CA | CB | CD | DA | DB | DC | AR1 | AR2 |
AR3 | AR4 | BR1 | BR2 | BR3 | BR4 | CR1 | CR2 | CR3 | CR4 |
DR1 | DR2 | DR3 | DR4 | R1A | R1B | R1C | R1D | R2A | R2B |
R2C | R2D | R3A | R3B | R3C | R3D | R4A | R4B | R4C | R4D |
R1R2 | R1R3 | R1R4 | R2R1 | R2R3 | R2R4 | R3R1 | R3R2 | R3R4 |
R4R1 | R4R2 | R4R3
Measurement parameter; see list of parameters below.
<TestPortNum>
Test port number, drive port for wave quantities and ratios, ignored
for S-parameters.
Range:
1 to 4 (depending on instrument model)
R&S ZVR/ZVAB Compatible Commands