GUI Reference
R&S
®
ZNC
186
User Manual 1173.9557.02 ─ 13
Note:
Ripple limit check and display of the ripple limit lines are independent of each
other: With disabled limit check, the limit line can still be displayed. If no limit lines are
defined for the active trace, the limit check can still be switched on but the info field will
display a warning ("No ripple defined!").
Remote command:
Ripple Fail Beep
Activates or deactivates the fail beep. The fail beep is a low-tone acoustic signal that is
generated each time the analyzer detects an exceeded ripple limit. No fail beep can be
generated if the ripple limit check is switched off.
Remote command:
Show Results All Traces
Shows or hides the info fields for all traces in the active recall set for which a limit check
is enabled, irrespective of the active trace. If "Show Results All Traces" is disabled (and
the ripple check is on), the info field for the active trace is displayed only.
Remote command:
No command; display configuration only.
Clear Test
Resets the limit check results.
Remote command:
4.2.5.4
Define Ripple Test (Dialog)
The "Define Ripple Test" dialog defines the ripple limits for the active trace on a range-
by-range basis. A separate ripple limit can be assigned to each range.
Trace Settings