Command Reference
R&S
®
ZNC
379
User Manual 1173.9557.02 ─ 13
<RvCount>
Number of sweep to be read. 1 denotes the last sweep acquired,
2 denotes the second-last and so forth.
Range:
1 to sweep count defined via
[SENSe<Ch>:]SWEep:COUNt
Example:
SWE:COUN 10
Define the number of sweeps (10) to be measured in single sweep
mode.
INIT:CONT OFF; :INIT; *OPC?
Activate single sweep mode and start a single sweep sequence in
channel no. 1. Wait until the single sweep sequence is complete.
CALC:DATA:NSW? SDAT,3
Query the results of the 8
th
sweep.
chapter 7.2.4.3, "Retrieving the Results of Previous
Usage:
Query only
CALCulate<Ch>:DATA:SGRoup?
<Format>
Reads the current response values of all S-parameters associated to a group of logical
ports (S-parameter group). The S-parameter group must be created before using
.
Suffix:
<Ch>
.
Channel number of the previously defined S-parameter group.
Query parameters:
<Format>
FDATa | SDATa | MDATa
Output format for the S-parameter data, see
Example:
See
Usage:
Query only
CALCulate<Chn>:DATA:STIMulus?
Reads the stimulus values of the active data or memory trace.
Suffix:
<Chn>
.
Channel number used to identify the active trace.
Example:
See
Usage:
Query only
CALCulate:DATA:TRACe?
<TraceName>, <Format>
Queries the trace data of an arbitrary (not necessarily the active) trace, referenced by its
trace name <TraceName>.
SCPI Command Reference