GUI Reference
R&S
®
ZNC
185
User Manual 1173.9557.02 ─ 13
Some of the softkeys are coupled to the softkeys in the "Limit Test" tab and provide
identical results:
●
●
Show Ripple Limits
Shows or hides the ripple limit lines associated with the active trace in a Cartesian dia-
gram area. The vertical positions of the ripple lines are re-calculated after each sweep;
only their stimulus range and distance (the ripple limit) is fixed.
Note:
Display of the ripple limit line and limit check are independent of each other: Hiding the
limit line does not switch off the limit check.
Remote command:
Ripple Check
Switches the ripple limit check of the active trace on or off.
When the limit check is switched on, a movable info field shows the pass/fail information
and the measured ripple in each ripple limit range. If the ripple limit check fails at a mea-
surement point, the point is marked with a colored square. The "Limit Fail Trace" color is
defined in the "Define User Color Scheme" dialog ("Display > Display Config. > Define
User Color..."). An acoustic signal ("Global Beep") and a TTL signal indicating pass or
fail can be generated in addition.
Trace Settings