Command Reference
R&S
®
ZNC
425
User Manual 1173.9557.02 ─ 13
Manual operation:
See
CALCulate<Ch>:PARameter:DELete
<TraceName>
Deletes a trace with a specified trace name and channel.
Suffix:
<Ch>
.
Channel number.
Setting parameters:
<TraceName>
Trace name, e.g. 'Trc4'. See "Rules for trace names" in
ter 4.2.4.2, "Trace Manager (Dialog)"
Example:
CALCulate4:PARameter:SDEFine 'Ch4Tr1', 'S11'
Create channel 4 and a trace named
Ch4Tr1
to measure the input
reflection coefficient S
11
.
CALCulate4:PAR:CAT?
Query the traces assigned to channel 4.
Ch4Tr1
is the only trace
assigned to channel 4, so the response is
'CH4TR1,S11'
.
CALCulate4:PARameter:SDEFine 'CH4TR2', 'S21';
SDEFine 'CH4TR3', 'S12'; SDEFine 'CH4TR4',
'S22'
Create three more traces for the remaining 2-port S-parameters.
CALCulate4:PARameter:DELete 'CH4TR1'
Delete the first created trace.
CALCulate4:PARameter:DELete:CALL
Delete the remaining three traces in channel 4.
CALCulate:PARameter:DELete:ALL
Delete all traces, including the default trace
Trc1
in channel 1.
Usage:
Setting only
Manual operation:
See
CALCulate:PARameter:DELete:ALL
Deletes all traces in all channels of the active recall set, including the default trace
Trc1
in channel 1. The manual control screen shows "No Trace".
Example:
See
Usage:
Event
Manual operation:
See
CALCulate<Ch>:PARameter:DELete:CALL
Deletes all traces in channel no.
<Ch>
.
Suffix:
<Ch>
.
Channel number
Example:
See
SCPI Command Reference