Currently, "non-standard" logical port assignments created via
can neither be created nor edited with the balanced port dialog. They
are erroneously displayed as the 1:1 standard mapping between physical and (single-
ended) logical ports.
Measurements do not necessarily require all of the physical and logical ports of the net-
work analyzer. To save measurement time, it is recommended to specify the ports that
are not used in the current test setup as "Unused" ports. Unused ports will not be con-
sidered for the calculation of mixed mode, Z- and Y-parameters.
Background information
Refer to the following sections:
●
chapter 3.3.5, "Unbalance-Balance Conversion"
●
chapter 3.3.5.1, "Balanced Port Configurations"
●
chapter 3.3.5.4, "Reference Impedances"
Predefined Configs
The "Predefined Configs" tab of the "Balanced and Measured Ports" dialog provides the
most commonly used balanced port configurations for the analyzer.
Access:
E.g. TRACE > MEAS > S-Params > Balanced Ports...
Trace Settings