GUI Reference
R&S
®
ZNC
225
User Manual 1173.9557.02 ─ 13
Seg x-Axis
Scales the x-axis for a segmented frequency sweep:
●
In frequency based mode, the x-axis covers the frequency ranges of all sweep seg-
ments, including possible gaps between the segments. Equal frequency spacings
correspond to equal distances on the x-axis.
●
In point based mode, the x-axis shows all sweep points with equal spacings. Gaps
between sweep segments are minimized; no diagram space is "wasted" on unused
frequency ranges. Point based mode is indicated in the channel line.
The example below shows a segmented frequency sweep with two segments. The first
segment ranges from 1 GHz to 1.4 GHz; the second segment from 2 GHz to 3 GHz. Both
segments contain 11 sweep points. In point based mode (lower diagram), all sweep
points are equidistant.
Tip:
Overlapping limit line and ripple limit line segments are not displayed when a point-
based x-xis is active; see
chapter 3.4.1.1, "Rules for Limit Line Definition"
Remote command:
4.4.2.3
Define Segments (Dialog)
The "Define Segments" dialog defines all channel settings for a "Segmented" frequency
sweep and imports or exports segmented sweep settings.
The dialog contains a table to edit the individual segments of the sweep range. Sweep
segments may have common points or even overlap.
See also
Access:
CHANNEL > SWEEP > Sweep Type > Define Segments...
Channel Settings