GUI Reference
R&S
®
ZNC
244
User Manual 1173.9557.02 ─ 13
Fig. 4-3: Calibration Presetting: Connectors and Cal Kits
The upper part of the panel shows the port(s) and the calibration type selected in
. The lower part gives access to the connector and cal kit settings.
Connector / Gender
Defines the connector types and genders of the ports to be calibrated. For symmetric
(sexless) connector types (e.g. 7 mm / PC7), "Gender" is unavailable.
If "Same Connector All Ports" is active, the connector types at all ports (but not their
gender) are always adjusted to the current selection. If "Same Gender All Ports" is active,
the genders at all ports are always adjusted to the current selection.
User-defined connectors can be added or removed in the
.
Remote command:
Cal Kit
Selects a cal kit for the connector at each selected physical port. The drop-down list
contains all available calibration kits for the selected connector type. The assignment of
a calibration kit to a connector type must be the same for all physical ports: If a calibration
kit is changed, the R&S ZNC automatically assigns the new kit to all ports with the same
connector type.
Channel Settings