GUI Reference
R&S
®
ZNC
241
User Manual 1173.9557.02 ─ 13
Background information
Refer to
, on page 67 for background information.
●
If the active channel is already equipped with a system error correction, the "Cali-
bration Presetting" wizard loads the underlying setup. If the calibration setup is not
changed and sweep data are available (see
existing system error correction can be optimized without repeating the measurement
of all standards.
●
When you apply the acquired system error correction, the active calibration is
replaced and discarded.
To persist any kind of calibration you can transfer it to the "Cal Pool" using the
.
●
The active system error correction data can be read (and modified) using the remote
control command
Step 1: Ports and Type
Allows to select the test ports to be calibrated and the calibration type to be used
Background information
Refer to
chapter 3.5.1, "Calibration Types"
Access:
CHANNEL > CAL > Start Cal > Start... (Manual)
Channel Settings