Command Reference
R&S
®
ZNC
431
User Manual 1173.9557.02 ─ 13
Setting parameters:
<SweepType>
FLIN | FLOG | FSEG | FSINgle | TLIN | TLOG | PLIN | PLOG |
PSINgle
Keywords for the units of the stimulus values; frequency, power,
and time units. The selected unit must be compatible with the
sweep type (
wise the ripple limit lines cannot be displayed and no ripple limit
check is possible.
*RST:
FLIN
Default unit: Hz (for FLIN, FLOG, FSEG, and FSINgle), s (for TLIN
and TLOG), dBm (for PLIN, PLOG and PSINgle).
Example:
SWE:TYPE POW
Select a power sweep.
CALC:RIPP:CONT:DOM PLIN
Delete all existing ripple limit ranges and select level units for the
ripple limit of the active trace.
CALC:RIPP:DATA 1, -10, -5, 3
Define and enable a ripple limit range in the stimulus range
between -10 dBm and -5 dBm, assigning a ripple limit of 3 dB.
Usage:
Setting only
Manual operation:
See
"Add / Insert / Delete / Delete All / Align All"
CALCulate<Chn>:RIPPle:DATA
<RippleLimRange>...
Adds and enables/disables an arbitrary number of ripple limit ranges, assigning the stim-
ulus values and the ripple limits. See
chapter 3.4.1.2, "Rules for Ripple Test Definition"
Note:
This command does not overwrite existing ripple limit ranges. The defined ranges
are appended to the range list as new ranges. Use the
CALCulate<Chn>:RIPPle:SEGMent<Seg>...
commands to change existing ripple
limits.
Suffix:
<Chn>
.
Channel number used to identify the active trace.
SCPI Command Reference