GUI Reference
R&S
®
ZNC
258
User Manual 1173.9557.02 ─ 13
ibration kit file, instead of the gender. During the calibration, the analyzer checks
whether the cal kit contains the necessary standard types for the required ports.
Remote command:
Add / Copy / Delete / View / Modify...
The buttons in the right part of the dialog are used to manage standards:
●
"Add" adds a new standard to the calibration kit. The properties of the standard can
be edited in the table.
●
"Copy" creates a new standard based on the properties of an existing standard.
●
"Delete" deletes the selected standard.
●
"View / Modify ..." opens the "View / Modify Cal Kit Standards" dialog. This dialog
shows the circuit model for the selected standard. For user-defined standard, you
can modify the circuit model parameters. See
chapter 4.4.3.9, "View / Modify Cal Kit
Remote command:
Read .s<n>p File...
Opens a file selection dialog where you can select a Touchstone file containing the
reflection or transmission S-parameters for the standard. The R&S ZNC uses the impor-
ted S-parameters rather than the circuit model to characterize the standard, if ".s<n>p
File" is checked in the standard table. The appropriate file type (*.s1p for one-port stand-
ards and *.s2p for two-port standards) is selected automatically.
Remote command:
4.4.3.9
View / Modify Cal Kit Standards Dialog
The "View / Modify Cal Kit Standards" dialog shows the circuit model for a selected cal-
ibration standard. It is also used to define or edit the circuit model (offset and load)
parameters for a user-defined standard.
Related information
Refer to the following sections:
●
chapter 3.5.2.4, "Cal Kit Files"
●
chapter 4.4.3.8, "Kit Standards Dialog"
●
chapter 3.5.2.1, "Calibration Standard Types"
Access:
CHANNEL > CAL > Calibration > Cal Kits ... > Standards ... > View / Modify
Channel Settings