GUI Reference
R&S
®
ZNC
195
User Manual 1173.9557.02 ─ 13
Marker Name
Assigns a (new) name to the active marker. Marker names may exceed the length of the
input box and contain letters, numbers, blanks and special characters.
Remote command:
Marker Format
Selects an output format for the (complex) active marker value in the movable marker
info field. The default marker format is the format of the associated trace.
All marker formats are always available, irrespective of the measured quantity. The output
values are calculated by a simple conversion of a complex measurement result; the
marker format defines the conversion rules. This flexibility in the calculation must be kept
in mind when interpreting the results and physical units displayed. See also
ter 3.2.4.6, "Measured Quantities and Display Formats"
Short description of marker formats
The formats of the markers assigned to a trace are independent of each other and of the
trace format settings. The following table gives an overview on how a complex marker
value z = x + jy is converted.
Table 4-2: Marker formats
Marker Format
Description
Formula
Default (Trace)
Marker format identical with trace format
–
Lin Mag
Magnitude of z, unconverted
|z| = sqrt ( x
2
+ y
2
)
dB Mag
Magnitude of z in dB
|z| = sqrt ( x
2
+ y
2
) dB Mag(z) =
20 * log|z| dB
Phase
Phase of z
φ (z) = arctan (y/x)
Trace Settings