Command Reference
R&S
®
ZNC
632
User Manual 1173.9557.02 ─ 13
Example:
*RST; :SWE:POIN 10
Reset the instrument and reduce the number of sweep points to
10.
SOUR:POW:CORR:ACQ PORT,1
Perform a source power calibration using port 1 as a source port.
SOUR:POW:CORR:DATA? 'A1'
Query the correction values. The analyzer returns 10 comma-sep-
arated real numbers.
SOUR:POW:CORR:DATA 'A1', 1, 2, 3, 4, 5, -6, -7,
-8, -9, -0
Replace the correction values by ten (new) numbers.
Manual operation:
See
SOURce<Ch>:POWer<PhyPt>:CORRection:GENerator<Gen>[:STATe]
<Boolean>
Enables or disables the source power calibration for channel
<Ch>
and for an external
generator number
<Gen>
. The command is disabled unless a source power calibration
for the external generator has been performed (
). To enable or disable a source power calibration for an ana-
.
Suffix:
<Ch>
.
Calibrated channel number
<PhyPt>
Calibrated port number. This suffix is ignored; the generator is
selected via
<Gen>
.
<Gen>
Generator number
Parameters:
<Boolean>
Enables (ON) or disables (OFF) the source power calibration for
port number
<PhyPt>
.
*RST:
OFF
Example:
See
SOURce<Ch>:POWer<PhyPt>:CORRection:NREadings
<NoCalSweeps>
Sets a limit for the number of calibration sweeps in the source power calibration. The
command is valid for all channels, ports and external generators. Equivalent command:
.
Suffix:
<Ch>
.
Calibrated channel number. This parameter is ignored; the limit
applies to all sources.
<PhyPt>
Calibrated port number. This parameter is ignored; the limit
applies to all sources.
SCPI Command Reference