Command Reference
R&S
®
ZNC
559
User Manual 1173.9557.02 ─ 13
<Characterization>
Name and (possibly) directory of the characterization (cal kit file)
to be used for the automatic calibration (string parameter):
–
If an empty string (
' '
) is specified, the factory cal kit file stored
in the active calibration unit is used.
–
A VNA cal kit file name
*.calkit
without path denotes a spe-
cific cal kit file stored in the active calibration unit.
–
A VNA cal kit file name
*.calkit
with path denotes a specific
cal kit file stored in an arbitrary directory.
<TestPort1>
Test port number. For an n-port automatic calibration, n arbitrary
(not necessarily consecutive) port numbers must be specified. For
a one path two port calibration (OPTPort), the first port no. denotes
the source port (fully calibrated port).
<CalUnitPort1>
Port numbers of the cal unit. For an n-port automatic calibration,
n arbitrary (not necessarily consecutive) port numbers must be
specified. It is possible to combine arbitrary (not necessarily
matching) pairs of analyzer and cal unit ports.
<TestPort2>
Second test port number
*RST:
n/a
Example:
CORR:COLL:AUTO:PORT FNPort, '', 1, 2, 2, 4, 4,
1
Perform an automatic full 3-port calibration at test ports 1, 2, and
4 using the calibration unit's default calibration kit file and ports 2,
4, and 1 of the cal unit.
See also
Usage:
Setting only
Manual operation:
See
[SENSe<Ch>:]CORRection:COLLect:AUTO:POWer
[<TestPort>]
Initiates a receiver power calibration at the analyzer port
<Port>
. This is the second step
of an automatic SMARTerCal.
Note:
The power calibration must be performed "after" the system error correction; see
example below.
Suffix:
<Ch>
.
Channel number of the calibrated channel
Setting parameters:
<TestPort>
Test port number. The port number must match the port number
selected via
;
see example.
SCPI Command Reference