SYSTem:COMMunicate:AKAL:CONNection
<CalStandard>, <Port>[, <SecondPort>]
Connects the selected calibration standard to one or two ports of the active calibration
unit (see
Setting parameters:
<CalStandard>
THRough | OPEN | SHORt | MATCh
Connected one or two-port standard
<Port>
Port number of the calibration unit, for one and two-port standards.
<SecondPort>
Second port number of the calibration unit, for two-port standards
(THRough) only, omitted for the one-port standards.
Example:
SYST:COMM:AKAL:CONN THR, 1, 2
Connect a trough standard between ports 1 and 2 of the cal unit.
Usage:
Setting only
Manual operation:
See
SYSTem:COMMunicate:AKAL:MMEMory[:STATe]
<Boolean>
Shows or hides the memory of the active calibration unit (see
).
Setting parameters:
<Boolean>
ON - memory is shown in a separate drive.
OFF - memory is not shown.
*RST:
OFF
Example:
SYST:COMM:AKAL:MMEM ON
Show the memory of the active calibration unit.
Usage:
Setting only
Manual operation:
See
SYSTem:COMMunicate:GPIB[:SELF]:ADDRess
<GBIBAddress>
Sets the GPIB address of the analyzer.
SCPI Command Reference