GUI Reference
R&S
®
ZNC
130
User Manual 1173.9557.02 ─ 13
Background information
Refer to the following sections:
●
chapter 3.3.2, "Impedance Parameters"
●
chapter 3.3.3, "Admittance Parameters"
●
chapter 3.3.2.2, "Z-Parameters"
●
chapter 3.3.3.2, "Y-Parameters"
Access:
TRACE > MEAS key or Alt + Shift + A
Y/Z-Parameter
Selects an Y- or Z-parameter as a measured quantity for the active trace. For an n-port
vector network analyzer, the pull-down list provides the full set of n
2
Y- and Z-parameters.
Y- and Z-parameters are expressed as Y/Z
<out>< in>
, where <out> and <in> denote the
output and input port numbers of the DUT.
Remote command:
"<Trace_Name>", "Y11" |
"Z11" ...
Create new trace and select trace name and measurement parameter:
"<Trace_Name>", "Y11" |
"Z11" ...
Y11 / Y12 / Y21 / Y22
Selects the 2-port Y- parameters as a measured quantity for the active trace. The Y-
parameters describe the admittances of a DUT with output ports terminated in a short
circuit (V = 0).
The four 2-port Y-parameters can be interpreted as follows:
●
Y
11
is the input admittance, defined as the ratio of the current I
1
to the voltage V
1
,
measured at port 1 (forward measurement with output terminated in a short circuit,
V
2
= 0).
Trace Settings