Command Reference
R&S
®
ZNC
405
User Manual 1173.9557.02 ─ 13
DEFault
The format of the trace no.
<Chn>
(
MLOG
after
*RST
); see
CALCulate<Ch>:FORMat
.
MLINear
|z| = sqrt ( x2 + y2 )
MLOGarithmic
MDB (for R&S ZVR compatibility)
|z| = sqrt ( x2 + y2 )
dB Mag(z) = 20 * log|z| dB
PHASe
φ
(z) = arctan( Im(z) / Re(z) )
POLar
COMPlex (for R&S ZVR compatibility)
x, y (Real and Imag)
GDELay
Group Delay,
–
d
Φ
(z) / d
Ω
REAL
x
IMAGinary
y
SWR
Standing Wave Ratio, SWR = (1 + |z|) / (1 |z|)
LINPhase
MLPhase (for R&S ZVR compatibility)
Lin Mag and Phase, |z|, arctan ( Im(z) / Re(z) )
LOGPhase
MDPhase (for R&S ZVR compatibility)
dB Mag and Phase, 20 * log|z| dB, arctan ( Im(z) / Re(z) )
IMPedance
R, X, L or C (depending on sign(X))
ADMittance
G, B, L or C (depending on sign(X))
CALCulate<Chn>:MARKer<Mk>:FUNCtion:BWIDth:MODE
<BandfilterType>
Selects the bandfilter search mode. In contrast to manual control, bandfilter tracking is
not automatically activated.
Suffix:
<Chn>
.
Channel number used to identify the active trace.
<Mk>
Marker number in the range 1 to 10. This numeric suffix is ignored
and may be set to any value because the bandfilter search func-
tions always use markers M1 to M4.
Parameters:
<BandfilterType>
BPASs | BSTop | BPRMarker | BSRMarker | BPABsolute |
BSABsolute | NONE
Bandfilter search type:
BPASs
–
Bandpass Search Ref to Max
BSTop
–
Bandstop Search Ref to Max
BPRMarker
–
Bandpass Search Ref to Marker
BSRMarker
–
Bandstop Search Ref to Marker
BPABsolute
–
Bandpass Absolute Level
BSABsolute
–
Bandstop Absolute Level
NONE
–
deactivate bandfilter search, result off
*RST:
NONE
SCPI Command Reference