Command Reference
R&S
®
ZNC
428
User Manual 1173.9557.02 ─ 13
Example:
CALC4:PAR:SDEF 'Ch4Tr1', 'S11'
Create channel 4 and a trace named Ch4Tr1 to measure the input
reflection coefficient S
11
.
DISP:WIND2:STAT ON
Create diagram area no. 2.
DISP:WIND2:TRAC:FEED 'CH4TR1'
Display the generated trace in diagram area no. 2.
Usage:
Setting only
Manual operation:
See
The measurement parameter is selected by means of the keywords in the following table.
The selection depends on the number of test ports of the analyzer, e.g. S
44
is not available
on 2-port analyzers.
Table 6-4: Name strings for measurement results
'S11' | 'S12' | ... | 'S0101' | ...
Single-ended S-parameters S<out><in>, where <out> and <in> denote the
numbers of the related (logical) output and input ports. To avoid ambiguities,
<out> and <in> must be either both 1-digit numbers (e.g. 12) or both 2-digit
numbers (e.g. 0102).
'SCD11' |
S-parameters involving balanced ports must be specified in the form
S<m_out><m_in><out><in>, where <m_out> and <m_in> denote the port
modes of the related (logical) output and input ports. All combinations of D
(differential, balanced), C (common, balanced) and S (single-ended, unbal-
anced) are allowed. Port numbers <out> and <in> like for normal mode S-
parameters.
'Y11' | ... | 'YSS11' | ... | 'YCC11' | ... | 'YDD11' | 'Z11'
| ... | 'ZSS11' | ... | 'ZCC11' | ... | 'ZDD11' | ...
Short-circuit Y-parameters and open-circuit Z-parameters with port modes and
port numbers like for normal mode S-parameters*).
'Y-S11' | ... | 'Y-SSS11' | ... | 'Y-SCC11' | ... | 'Y-SDD11'
| 'Z-S11' | ... | 'Z-SSS11' | ... | 'Z-SCC11' | ... | 'Z-
SDD11' | ...
S-parameters converted to matched-circuit admittances and impedances with
port modes and port numbers like for normal mode S-parameters.
'A1' | ... | 'A01' | ... | 'B1' | ... | 'B01' | ... | 'A1SAM' |
'A1AVG' | 'A1AMP'
Wave quantities with port numbers like for normal mode S-parameters. The
strings SAM, AVG, AMP appended to the wave quantities denote a normal
(sample, SAM), AVG Real Imag (AVG), or AVG Mag Phase (AMP) detector.
The observation time for average detectors is set via
[SENSe<Ch>:]SWEep:DETector:TIME
.
'A1D2' | ... | 'A01D02' | ... | 'B1D2' | ... | 'B01D02' | ... |
'A1D1SAM' | 'A1D1AVG' | 'A1D1AMP' | ...
Wave quantities with port numbers and source port numbers (D<no> for drive
port). The strings SAM, AVG, AMP appended to the wave quantities denote a
normal (sample, SAM), AVG Real Imag (AVG), or AVG Mag Phase (AMP)
detector.
'B2/A1' | ... 'B02/A01' | ... | 'B2/A1SAM' | 'B2/A1AVG'
| B2/A1AMP' | ...
Ratio of wave quantities with port numbers like for normal mode S-parameters.
The strings SAM, AVG, AMP appended to the wave quantities denote a normal
(sample, SAM), AVG Real Imag (AVG), or AVG Mag Phase (AMP) detector.
'B2D1/A1D1' | ... 'B02D01/A01D01' | ... 'B2D1/
A1D1SAM' | 'B2D1/A1D1AVG' 'B2D1/A1D1AMP' | ...
Ratio of wave quantities with port numbers and source port numbers (D<no>
for drive port). The strings SAM, AVG, AMP appended to the wave quantities
denote a normal (sample, SAM), AVG Real Imag (AVG), or AVG Mag Phase
(AMP) detector.
'IMB21' | 'IMB12' | 'IMB31' | ...
Imbalance parameter for balanced ports Imb<receive_port><drive_port>. The
indices denote the logical receive port and the logical drive port of the analyzer.
'KFAC21' | 'KFAC12' | ...
Stability factor K (for unbalanced ports only)
'MUF121' | 'MUF112' | ...
Stability factor 1 (for unbalanced ports only)
SCPI Command Reference