GUI Reference
R&S
®
ZNC
247
User Manual 1173.9557.02 ─ 13
●
For a TRL calibration, at least one line standard must be measured between any pair
of calibrated ports. See
chapter 3.5.1.8, "TRL Calibration"
Remote command:
See also:
Restart Sweep on Std Meas
If this function is active, a new standard measurement initiates a new sweep, starting at
the beginning ("Start") of the sweep range: The sweep points for the calibration sweep
are in ascending order, like for an ordinary measurement.
If "Restart Sweep on Std Meas" is not active, the new standard measurement is started
at the current sweep point; the current sweep is continued as a calibration sweep.
Apply
Is enabled as soon as sufficient data have been acquired for the calibrated ports and
standards. The button starts the calculation of the system error correction and closes the
calibration wizard. The current instrument settings is stored with the correction data.
To avoid incompatibilities, older calibration data is deleted unless it has been transferred
into a "Cal Pool" using the
.
Note: Checks during the calculation of correction data
Incompatibilities between the selected calibration type, the standards and the channel
settings may cause the calibration to be inaccurate. The analyzer auto-detects potential
sources of errors and displays appropriate, self-explanatory notice boxes.
Remote command:
4.4.3.4
Power Cal Screen
The "Power Cal" screen shows the current source ("Power") and receive ("Meas.
Receiver") ports and allows you to perform scalar source power calibrations (flatness
calibrations) or measurement receiver calibrations, based on the current power calibra-
tion settings. If the active recall set contains several channels, an "Info" box lets you
decide whether the R&S ZNC acquires calibration data for the active channel or for all
channels.
Access:
CHANNEL > CAL > Start Cal > Power Cal...
Channel Settings