Command Reference
R&S
®
ZNC
571
User Manual 1173.9557.02 ─ 13
[SENSe<Ch>:]CORRection:COLLect:SCONnection<PhyPt>
<Type>[, <Gender>]
Selects a connector type at a specified port
<PhyPt>
and its gender. In contrast to
, this command uses a
string variable to identify the connector type.
Suffix:
<Ch>
.
Channel number of the calibrated channel
<PhyPt>
Port numbers of the analyzer. If the analyzer is set to use the same
connectors at all ports (
ALL
), then a change of a con-
nector type is valid for all ports. The gender of the connectors can
still be different.
Parameters:
<Type>
Connector type (string parameter) of the connectors. See "Cal kit
naming conventions" under
*RST:
'N 50 Ohm',FEM for all ports.
<Gender>
MALE | FEMale
Gender of the connectors. The gender designation is not neces-
sary (and ignored) for sexless connector types.
Example:
*RST; :CORR:COLL:SCON1 'N 75 Ohm', MALE; SCON4?
Change the connector type at port 1 from 'N 50 Ohm', FEM to 'N
75 Ohm', MALE. The connector type at the other ports is also
changed to N 75 Ohm, however, the gender (female) is main-
tained.
CORR:COLL:SCON4?
returns
'N 75 Ohm', FEM
.
Manual operation:
See
[SENSe<Ch>:]CORRection:CONNection
<ConnectorName>[, <Mode>, <Gender>,
<RelPermittivity>, <RefImpedance>]
Configures the user-defined connector types.
Suffix:
<Ch>
.
Channel number
Parameters:
<ConnectorName>
Name of the user-defined connectors, string parameter.
<Mode>
TEM | WGUide
Transverse electric or waveguide type propagation mode.
<Gender>
GENDer | NGENder
GENDer
–
polar connector type (m/f)
NGENder
–
sexless connector type
<RelPermittivity>
Relative permittivity
Range:
0.0000000001 to 1000.
SCPI Command Reference