Command Reference
R&S
®
ZNC
419
User Manual 1173.9557.02 ─ 13
Example:
Suppose that the active recall set contains an active trace no. 1.
CALC:MARK ON
Create marker no. 1 and display it in the center of the sweep range.
CALC:MARK:Y?
Query the measurement value at the marker position.
CALC:MARK:STAT:AREA LEFT, TOP
For a subsequent check at the GUI or a hardcopy, move the info
field to the top left position.
Usage:
Query only
Manual operation:
See
6.3.1.9
CALCulate:MATH...
The
CALCulate:MATH...
commands permit processing of measured data in numerical
expression format. The operators are +, -, *, / and use of constants and data arrays are
permitted.
..............................................................419
..................................................................................420
.................................................................................421
......................................................................................421
..........................................................................422
CALCulate<Chn>:MATH[:EXPRession]:SDEFine
<Expression>
Defines a general mathematical relation between traces. To calculate and display the
new mathematical trace, the mathematical mode must be switched on (
ON
).
Suffix:
<Chn>
.
Channel number used to identify the active trace.
Parameters:
<Expression>
Operands, operators and functions; see table below.
Example:
*RST; :CALC:MATH:MEM
Copy the current state of the default trace
'Trc1'
to a memory
trace named
'Mem2[Trc1]'
. The memory trace is not displayed.
CALC:MATH:SDEF 'Trc1 / Mem2[Trc1]'
Define a mathematical trace, dividing the data trace by the stored
memory trace. The mathematical trace is not displayed.
CALC:MATH:STAT ON
Display the mathematical trace instead of the active data trace.
Manual operation:
See
Expressions defined via
CALCulate<Ch>:MATH[:EXPRession]:SDEFine
may con-
tain the following elements:
SCPI Command Reference