Command Reference
R&S
®
ZNC
599
User Manual 1173.9557.02 ─ 13
Suffix:
<Ch>
.
Channel number
Example:
SEGM:ADD
Create a new sweep segment no. 1 in channel no. 1 using default
settings.
SEGM OFF
Disable the measurement in the created sweep segment.
SEGM:COUN?
Query the number of segments.
Usage:
Query only
Manual operation:
See
[SENSe<Ch>:]SEGMent<Seg>:DEFine
<StartFreq>, <StopFreq>, <Points>, <Power>,
<SegmentTime>|<MeasDelay>, <Unused>, <MeasBandwidth>[, <LO>,
<Selectivity>]
Creates or re-defines a sweep segment no.
<Seg>
with specific channel settings.
Entry of the first seven numeric parameters is mandatory; no default values are provided.
All settings except
<LO>
can be changed for existing segments using other commands
of the
[SENSe<Ch>:]SEGMent<Seg>...
subsystem.
Note:
to create a segment with default chan-
(no query) to insert a new
segment into the current segment list.
Suffix:
<Ch>
.
Channel number
<Seg>
Sweep segment number. Segment numbers must be sequential.
The specified segment number must be smaller or equal to the
number of existing segments plus 1.
If segment number
<Seg>
already exists, it is replaced by the new
segment.
Parameters:
<StartFreq>
Start frequency of the segment; see
Default unit: Hz
<StopFreq>
Stop frequency of the segment; see
Default unit: Hz
<Points>
Number of sweep points in the segment. See
<Power>
Internal source power in the segment. See
Default unit: dBm
SCPI Command Reference