Command Reference
R&S
®
ZNC
603
User Manual 1173.9557.02 ─ 13
Example:
SEGM:INS 1MHZ, 1.5MHZ, 111, -21DBM, 0.5S, 0,
10KHZ
Create a sweep segment with a sweep range between 1.0 MHz
and 1.5 MHz.
SEGM:FREQ:STAR?; STOP?
Query the start and stop frequency of the created segment. The
response is
1000000;1500000
.
Manual operation:
See
Note:
If the start frequency entered is greater than the current stop frequency, the stop
frequency is set to the start frequency plus the minimum frequency span of 1 Hz. If the
stop frequency entered is smaller than the current start frequency, the start frequency is
set to the stop frequency minus the minimum frequency span (
).
[SENSe<Ch>:]SEGMent<Seg>:INSert
<StartFreq>, <StopFreq>, <Points>, <Power>,
<SegmentTime>|<MeasDelay>, <Unused>, <MeasBandwidth>[, <LO>,
<Selectivity>]
Adds a new sweep segment no.
<Seg>
with specific channel settings.
Entry of the first seven numeric parameters is mandatory; no default values are provided.
All settings except
<LO>
can be changed for existing segments using other commands
of the
[SENSe<Ch>:]SEGMent<Seg>...
subsystem.
Note:
to create a segment with default chan-
to change or query all set-
tings of an existing segment.
Suffix:
<Ch>
.
Channel number
<Seg>
Sweep segment number. Segment numbers must be sequential.
The specified segment number must be smaller or equal to the
number of existing segments plus 1.
If one or more sweep segments with segment numbers
<Seg>
or
larger exist in the current channel, then all these existing segment
numbers are incremented by 1 and the new segment is inserted
as segment no.
<Seg>
.
Parameters:
<StartFreq>
Start frequency of the segment. See
Default unit: Hz
<StopFreq>
Stop frequency of the segment; see
Default unit: Hz
<Points>
Number of sweep points in the segment. See
SCPI Command Reference