Command Reference
R&S
®
ZNC
610
User Manual 1173.9557.02 ─ 13
Example:
SEGM:ADD
Create a new sweep segment no. 1 in channel no. 1 using default
settings and thus 0 s meas. delay.
SEGM:SWE:DWELl 0.1
Increase the meas. delay to 0.1 s.
SEGM:SWE:DWELl:CONT OFF
Couple the meas. delay in all segments and reset the delay in
segment no. 1 to the initial value of 0 s.
Manual operation:
See
[SENSe<Ch>:]SEGMent<Seg>:SWEep:POINts
<SegPoint>
Defines the total number of measurement points in sweep segment no.
<Seg>
.
Suffix:
<Ch>
.
Channel number
<Seg>
Sweep segment number
Parameters:
<SegPoint>
Number of points in the segment
Range:
1 to 100001. 1 is allowed if start and stop frequencies
are equal.
*RST:
51
Example:
SEGM:ADD
Create a new sweep segment no. 1 in channel no. 1 using default
settings.
SEGM:SWE:POIN 401
Increase the number of points to 401.
Manual operation:
See
[SENSe<Ch>:]SEGMent<Seg>:SWEep:TIME
<SegSweepTime>
Sets the duration of the sweep in sweep segment no.
<Seg>
("Segment Time"). At the
same time, the command activates separate sweep time setting in all sweep segments
(
ON
).
Suffix:
<Ch>
.
Channel number
<Seg>
Sweep segment number
SCPI Command Reference