Table Columns
The table in the upper part of the "Define Segments" dialog contains an automatically
assigned current number for each segment plus the following editable or non-editable
columns:
●
"On" provides check boxes to activate or deactivate each individual segment.
"Sweep" points in inactive segments are not measured and not listed in the "Point
List".
●
"Start" is the stimulus (x-axis) value of the first point of the segment. If the segment
contains more than one "Point", then "Start" must be smaller than the "Stop" value.
If a "Start" value equal to or larger than the current "Stop" value is set, "Stop" is
adjusted to the new "Start" value plus 1 Hz.
●
"Stop" is the stimulus (x-axis) value of the last point of the segment. If the segment
contains more than one "Point", then "Stop" must be larger or equal than the "Start"
value. If a "Stop" value equal to or smaller than the current "Start" value is set,
"Start" is adjusted to the new "Stop" value minus 1 Hz.
●
"Points" is the number of sweep points in the segment. A single segment can consist
of only one point. If "Points" is set to 1, then the "Stop" frequency is set equal to the
"Start" frequency.
●
The remaining columns show the channel settings for each segment. They are dis-
played only if they are selected in the "Displayed Columns" dialog.
Note: Limitations for overlapping segments
Channel Settings