User Manual 1173.9557.02 ─ 13
Parameters:
<SegSweepTime>
Segment time. The minimum segment time depends on the other
channel settings, in particular on the number of points (
) and the delay for each partial measurement (
). The maximum
is 1000 s. Changing the duration leaves the number of points
unchanged but directly affects the delay.
Range:
Minimum value to 1000 s
*RST:
Minimum value, depending on the channel settings.
This default value corresponds to automatic sweep
time setting in manual control.
Default unit: s
Example:
SEGM:ADD
Create a new sweep segment no. 1 in channel no. 1.
SEGM:SWE:TIME 0.1
Increase the segment sweep time to 0.1 s.
SEGM:SWE:TIME:SUM?
Query the total duration of the segmented sweep. The response
is 0.1, because there is only one sweep segment.
Manual operation:
See
[SENSe<Ch>:]SEGMent<Seg>:SWEep:TIME:CONTrol
<Boolean>
Selects common or independent "Segment Sweep Time" settings for the sweep seg-
ments.
Suffix:
<Ch>
.
Channel number
<Seg>
Sweep segment number. This suffix is ignored; the command
controls the whole segmented sweep.
Parameters:
<Boolean>
ON
–
use independent segment sweep time settings, to be defined
.
OFF
–
reset the segment sweep time in all sweep segments to the
segment sweep time for unsegmented sweeps, defined via
. ON will not restore the previous
values.
The parameter is automatically switched to
ON
when a segment
or if the channel settings in a sweep segment
require a sweep time larger than the unsegmented sweep time.
*RST:
OFF
SCPI Command Reference