Command Reference
R&S
®
ZNC
465
User Manual 1173.9557.02 ─ 13
CONTrol:HANDler:LOGic
<Logic>
Selects the logic of the data ports A to H of the Universal Interface. For output ports, a
change in logic reverses the state of the output lines immediately. For input ports, a
change in logic will be reflected next time when data is read.
Parameters:
<Logic>
POSitive | NEGative
POSitive
–
0 = low, 1 = high
NEGative
–
0 = high, 1 = low
Note:
*RST
or "Preset" do not change the configuration of the Uni-
versal Interface. Use
default values.
*RST:
n/a (default: NEGative)
Example:
CONTrol:HANDler:LOGic POS
Change the logic of the data ports to positive.
CONTrol:HANDler:OUTPut<Pt>:USER
<BinValue>
Defines the state of the output ports (pin 3 or 4) of the Universal Interface connector after
the next negative pulse on the /INPUT1 line (pin 2).
Suffix:
<Pt>
.
Output port number:
1 - /OUTPUT1 (pin 3)
2 - /OUTPUT2 (pin 4)
The parameters
MIN
,
MAX
,
DEF
are not supported.
Note:
*RST
or "Preset" do not change the configuration of the Uni-
versal Interface. Use
default values.
Parameters:
<BinValue>
0 - high 1 - low
*RST:
n/a (default: 0)
Example:
See
CONTrol:HANDler:OUTPut<Pt>[:DATA]
<BinValue>
Writes a 0 or 1 to the output ports (pin 3 or 4) of the Universal Interface connector. The
port lines have negative logic: A "0" corresponds to a high signal, a "1" to a low signal.
The query reads the last value that has been written to the output port.