GUI Reference
R&S
®
ZNC
238
User Manual 1173.9557.02 ─ 13
If auto-detection fails because of a high attenuation in the signal path, you can either
enter the port assignment manually or connect matching port numbers and select "Default
Port Assignment".
Remote command:
Default Port Assignment
Restores the default port assignment.
Remote command:
Start
If the configured port assignments are invalid, this action is disabled.
Step 3: Calibration
During the calibration phase the R&S ZNC displays a "Cal Unit" screen that guides the
user through the actual correction data acquisition, whose setup was prepared during the
previous wizard steps.
In each step
●
the calibration unit has to be (re-)connected according to the depicted port assign-
ment; auto-detection is possible
●
the related test ports are calibrated with the same calibration type that was selected
for in the
●
a subsweep is performed for every required test port (pair), for every possible path
(if external switch matrices are involved) and for every required standard
When these steps have been completed, the resulting system error correction can be
calculated and applied.
Channel Settings