GUI Reference
R&S
®
ZNC
242
User Manual 1173.9557.02 ─ 13
Fig. 4-2: Calibration Presetting: Ports and Type
Ports
Selects the test port(s) to be calibrated.
Remote command:
The port parameters in many calibration commands define the calibrated port(s).
Type / Source
Selects the calibration type. The green arrow symbols give a preview of the type and the
number of calibration sweeps involved:
●
Curved arrows (example: "Refl Norm Open") denote one or more reflection meas-
urements at each port.
●
Straight, horizontal arrows (example: "Trans Norm") denote one or more transmission
measurements between each pair of two ports.
●
The full n-port calibration types (n > 1, e.g. TOSM, TRL...) are symbolized by a closed
square symbol. The number of arrows increases the complexity but can also improve
the accuracy of the calibration.
The reflection calibration types can be used for any set of test ports: reflection calibrations
are simply repeated for each port. A transmission calibration type requires at least two
physical ports. For the unidirectional transmission calibration types ("Trans Norm", "One
Path Two Ports"), the (
"Source"
port) must be specified in addition. For an overview refer
to
.
Remote command:
Channel Settings