Test Port Assignment
Defines the assignment between test ports and cal unit ports. In the default "Manual"
assignment, VNA ports and cal unit port numbers match. If you decide to use a different
assignment, you can auto-detect the actual assignment ("Automatic") or select the ana-
lyzer port numbers manually. Auto-detection may fail e.g. because of a high attenuation
in the signal path.
Remote command:
Take OSM ... / Take All OSM and Through
Starts a calibration sweep for the selected port(s). "Take All OSM and Through" initiates
a series of calibration sweeps; the R&S ZNC acquires a full set of one-port and two-port
data. This is required for the transmission normalizations and for a "One Path Two
Ports" calibration; see
"Dependency between calibration types and characterization
Remote command:
Next
Proceeds to the second dialog in the characterization wizard ("Save Characterization
Data"). Next is available as soon as the R&S ZNC has acquired characterization data for
a single port.