1. Features 

 

• 

3 ½-digit LCD display with low-battery indication 

• 

one single operating switch 

• 

light pocket-sized model 

• 

measures luminous intensity from 0.01 to 50.000lux 

• 

2000lux : reading x 10 // 50.000lux : reading x 100 

 

2. Specifications 

 
Display 

:  1999 counts & low-battery indication 

Low-Battery Indication 

:  “BAT” is displayed when the battery voltage drops below the operating level 

Measurement Rate 

:  1.5 readings/sec., nominal 

Storage Temperature 

:  -10 to +60°C / 14 to 140°F with relative humidity < 80% 

Power 

:  1 x standard A23 battery of 12V 

Dimensions Photo Detector  :  115 x 60 x 27mm 

Dimensions Device 

:  188 x 64.5 x 24.5mm 

Weight : 

160g 

 
Measuring Range 

:  200, 2000, 20000lux (reading x 10) and 50000lux (reading x 100) 

Overrange Indication 

:   “1” is displayed 

Measuring Accuracy 

:  ±5% of rdg + 10 digits (<10000lux) 

 

±10% of rdg + 10 digits (>10000lux) 

Precision : 

±0.1%/°C 

Photo Detector 

:1 silicon photo diode with filter 

 

3. Front Panel Description 

 
1.  LCD display 

:  3 ½-digit LCD display with low-battery indication (“BAT”) 

2.  Power/Function/range switch  :  Turns the power ON/OFF and selects the measurement function and the range 
3.  Photo detector 

:  Long-life silicon photo diode is built in 

 

4. Operating Instructions 

 
1.  Select the desired scale with the operating switch and select the desired range (“lux”, “lux x 10” or “lux x 100”). 
2.  Hold the photodetector at a horizontal angle to the light source to be measured. 
3.  Read the nominal measured value from the LCD display. 

4.  Overrange indication : if the instrument only display one “1” in the LCD, the input signal is too strong for the 

selected range. 

5.  Replace the photo detector when the measurement is completed. 
6.  The spectral sensitivity of this device is very similar to the ICE curve (International Commission on Illumination) 

depicted below on page 2 : 

Содержание DVM1300

Страница 1: ...of rdg 10 digits 10000lux 10 of rdg 10 digits 10000lux Precision 0 1 C Photo Detector 1 silicon photo diode with filter 3 Front Panel Description 1 LCD display 3 digit LCD display with low battery ind...

Страница 2: ...ans un h tel public room cloakroom vergaderruimte vestiaire salle de r union vestiaire 100 200 reception receptie r ception 200 1000 Store In een winkel Dans un magasin staircase trappenhuis cage d es...

Страница 3: ...BAT verschijnt wanneer de batterijspanning onvoldoende is om het toestel te kunnen gebruiken Bemonsteringsfrequentie 1 5 uitlezingen sec nominaal Opslagtemperatuur 10 tot 60 C 14 tot 140 F met relatie...

Страница 4: ...eiding kunnen worden gewijzigd zonder voorafgaande kennisgeving DVM1300 LUXMETRE 1 Caract ristiques LCD 3 digits avec indication pile faible Op ration avec 1 seul commutateur Mod le l ger en format de...

Страница 5: ...il ouvrez le compartiment de la pile et remplacez la A23 de 12V Refermez le compartiment Remarque le contenu et les sp cifications du manuel peuvent tre modifi es sans notification pr alable DVM1300 L...

Страница 6: ...la de la curva ICI Comisi n internacional de iluminaci n v ase abajo 7 Iluminaci n recomendada v ase la siguiente lista ENGLISH ESPA OL DEUTSCH LUX At the Office En la oficina B ro conference room re...

Страница 7: ...A23 de 12V Cierre el compartimiento Observaci n Se pueden modificar las especificaciones y el contenido de este manual sin previo aviso DVM1300 LUXMETER 1 Eigenschaften 3 stelliges LCD Display mit Low...

Страница 8: ...e den im Display angezeigten nominellen ab 4 berlaufanzeige Bei Anzeige 1 ist der gew hlte Messbereich berlastet 5 Setzen Sie den Lichtsensor zur ck nach Abschluss der Messung 6 Die spektrale Empfindl...

Отзывы: