Command Reference
R&S
®
ZNC
464
User Manual 1173.9557.02 ─ 13
CONTrol:HANDler[:EXTension]:INDex:STATe
<Boolean>
Selects the digital signal that is routed to pin 20 of the Universal Interface connector.
Parameters:
<Boolean>
ON - /INDEX signal at pin 20
OFF - /PORT_B6 signal at pin 20
*RST:
n/a (default: OFF)
Note:
*RST
or "Preset" do not change the configuration of the Uni-
versal Interface. Use
default values.
Example:
CONT:HAND:EXT:IND:STAT ON
Route the /INDEX signal to pin 20.
CONT:HAND:EXT:RTR:STAT ON
Route the /READY_FOR_TRIGGER signal to pin 21.
CONT:HAND:RES
Restore the default state: Pins no. 20 and 21 are available for port
B input/output signals.
CONTrol:HANDler[:EXTension]:RTRigger:STATe
<Boolean>
Selects the digital signal that is routed to pin 21 of the Universal Interface connector.
Parameters:
<Boolean>
ON - /READY_FOR_TRIGGER
OFF - /PORT_B7 signal at pin 21
Note:
*RST
or "Preset" do not change the configuration of the Uni-
versal Interface. Use
default values.
*RST:
n/a (default: OFF)
Example:
See
CONTrol:HANDler:INPut?
<NumberOfTrans>
Queries whether a high to low transition occurred at the /INPUT 1 line (pin 2) of the
Universal Interface since the last
CONTrol:HANDler:INPut?
query. A negative pulse
fed to this line also causes the /OUTPUT 1 and /OUTPUT 2 lines (pins 3 and 4) to change
to low.
Parameters:
<NumberOfTrans>
0
–
no transition detected since last query.
1
–
one or more transitions detected. The query resets the counter
to zero.
*RST:
n/a
Example:
CONTrol:HANDler:INPut?
Query whether a high to low transition occurred.
Usage:
Query only
SCPI Command Reference