Command Reference
R&S
®
ZNC
515
User Manual 1173.9557.02 ─ 13
<CalGroupFile>
Optional string parameter to specify the name of the cal group file.
Cal group files must have the extension
*.cal
. The directory must
not be specified; the analyzer always uses the default path
C:
\Users\Public\Documents\Rohde-Schwarz\Vna
\Calibration\Data
. If there is no link between
<Ch>
and the
specified file, the command has no effect. If no file is specified, the
command resolves any link between
<Ch>
and an arbitrary cal
group file.
Example:
See
Usage:
Setting only
Manual operation:
See
MMEMory:LOAD:CORRection:TCOefficient
<TraceFile>[, <Trace>]
Loads trace data from a specified power meter correction file or trace file and assigns it
to a trace with a specified name. Traces are created using
.
Setting parameters:
<TraceFile>
String parameter to specify the name and directory of the loaded
trace file. The R&S ZNC supports power meter correction list files
(
*.pmcl
, to be generated using
*csv
, and Touchstone
(
*.s<n>p
) files. If no path is specified the analyzer searches the
current directory, to be queried with
?
.
The file extensions
*.s<n>p
,
*csv
, and
*.pmcl
for Touchstone,
ASCII, and power meter correction list files are mandatory.
*RST:
n/a
<Trace>
Optional string parameter: For inported Touchstone files for more
than one port (
*.s2p
,
*.s3p
,
*.s4p
), the parameter denotes the
imported S-parameter (
'S11', 'S12', ...
). For ASCII
(
*.csv
) files, the parameter references a trace name in the file
(case sensitive). If the parameter is omitted, the first trace in the
specified file is imported.
The parameter is not used for power meter correction list files
(*.pmcl).
*RST:
n/a
Example:
See
Usage:
Setting only
Manual operation:
See
SCPI Command Reference