Parameters:
<Activate>
ON | OFF - switch the display on or off. If the display is switched
on, the analyzer shows the diagrams and traces like in manual
control.
ONCE - switch the display on and show the current trace. This
parameter can be used for occasional checks of the measurement
results or settings. The measurement is continued, however, the
measurement results are not updated. Compared to the
ON
set-
ting,
ONCE
does not slow down the measurement speed.
Example:
SYST:DISP:UPD ON
Switch the display on to view the traces and diagrams.
SYSTem:ERRor:ALL?
Queries and at the same time deletes all entries in the error queue.
The entries consist of an error number and a short description of the error. Positive error
numbers are instrument-dependent. Negative error numbers are reserved by the SCPI
standard; see section
Error Messages and Troubleshooting
Example:
SYST:ERR:ALL?
Query all entries in the error queue.
0,"No error"
is returned if
the error queue is empty.
Usage:
Query only
SYSTem:ERRor:DISPlay
<Boolean>
Switches the display of a tooltip for remote command errors on or off. The tooltip appears
at the bottom of the remote screen and of the manual screen; it is
not
displayed for SCPI
errors no.
-113, Undefined header
.
Parameters:
<Boolean>
ON | OFF - switch the tooltip on or off.
*RST:
OFF
Example:
SYST:ERR:DISP ON
Switch the display of the tooltip for remote command errors on.
FREQ:STAR 1
Generate a
Remote Error: -222, "Data out of
range; ..."
tooltip.
Manual operation:
See
SYSTem:ERRor[:NEXT]?
Queries and at the same time deletes the oldest entry in the error queue. Operation is
identical to that of