Remote Control
R&S
®
ZNC
333
User Manual 1173.9557.02 ─ 13
It remote logging is enabled (
SYSTem:LOGGing:REMote[:STATe] ON
) the analyzer
stores all received commands to the file
'C:\Users\Public\Documents\Rohde-Schwarz\Vna\RemoteLog'
.
5.1.3 Switchover to Remote Control
On power-up, the instrument is always in the manual operating state and can be operated
via the front panel controls. The instrument is switched to remote control as soon as it
receives a command from the controller. If the instrument is controlled via RSIB or VXI-11
protocol, the alternative commands @REM and @LOC can be used to switch from man-
ual to remote control and back.
While remote control is active, operation via the front panel is disabled with the exception
of the "Remote" softtool panel. The instrument settings are optimized for maximum mea-
surement speed; the display is switched off:
The softkeys in the remote screen are used to modify or quit the remote state:
●
"Go to Local" switches the instrument to local state.
●
"Display" switches the display on or off.
●
If a remote error message is displayed at the bottom of the remote screen, you can
use "Clear Error Messages" to delete it.
The remaining controls are for future extensions.
Display on and off states
Switching on the display means that the analyzer shows the measurement screen with
the current recall sets, diagram areas and traces without leaving the remote state. In this
operating mode, it is possible to observe the screen while a remote control script is exe-
cuted and the control elements on the front panel are still disabled.
Introduction to Remote Control