Command Reference
R&S
®
ZNC
579
User Manual 1173.9557.02 ─ 13
Example:
See
Manual operation:
See
"Loss at DC / Loss at Freq / Freq for Loss"
[SENSe<Ch>:]CORRection:LOSS<PhyPt>:AUTO
<Activate>
Defines the offset parameters for the active test port such that the residual delay of the
active trace (defined as the negative derivative of the phase response) is minimized and
the measured loss is reproduced as far as possible across the entire sweep range ("Auto
Length and Loss").
Suffix:
<Ch>
.
Channel number of the offset-corrected channel
<PhyPt>
Port number of the analyzer. This numeric suffix is ignored; the
active port is determined by the active trace.
Setting parameters:
<Activate>
ONCE
Applies the "Auto Length and Loss" function.
Example:
*RST; :CORR:LOSS:AUTO ONCE
Reset the instrument and apply the "Auto Length and Loss" func-
tion to the default trace (Trc1 in channel 1).
Usage:
Setting only
Manual operation:
See
[SENSe<Ch>:]CORRection:LOSS<PhyPt>:FREQuency
<RefFreq>
Defines the reference frequency for the frequency-dependent part of the offset loss (
).
Suffix:
<Ch>
.
Channel number of the offset-corrected channel
<PhyPt>
Port number of the analyzer
Parameters:
<RefFreq>
Reference frequency
Range:
Frequency range of the analyzer model.
Increment: 1 MHz
*RST:
1000000000 Hz (= 1 GHz)
Default unit: Hz
Example:
See
Manual operation:
See
"Loss at DC / Loss at Freq / Freq for Loss"
[SENSe<Ch>:]CORRection:LOSS<PhyPt>:OFFSet
<OffsetLoss>
Defines the frequency-independent part (DC value) of the offset loss.
SCPI Command Reference