GUI Reference
R&S
®
ZNC
294
User Manual 1173.9557.02 ─ 13
See also
chapter 3.6.1.2, "Definition of Loss Parameters "
Note:
The entered parameters define the loss for a signal traveling in one direction
through the transmission line. To account for the propagation in both directions, the mag-
nitude shift of a reflection parameter due to a given loss is twice the magnitude shift of a
transmission parameter. See also
chapter 3.6.1.6, "Application and Effect of Offset
Remote command:
Auto Length and Loss
Determines all offset parameters such that the residual group delay of the active trace
(defined as the negative derivative of the phase response) is minimized and the mea-
sured loss is minimized as far as possible across the entire sweep range.
See also
chapter 3.6.1.4, "Auto Length and Loss"
Note:
If "Auto Length and Loss" is used with a line connected to a test port, the end of
the line should be left open.
Remote command:
Reset Offsets
Resets the loss parameters to zero and the reference frequency to 1 GHz.
Remote command:
4.5 Display Settings
The "Display" menu provides all display settings and the functions for activating, modi-
fying and arranging different diagrams.
Related information
Refer to the following sections:
●
chapter 3.1.3, "Traces, Channels and Diagrams"
●
chapter 3.2.2, "Display Elements in the Diagram"
●
See chapter "Operating the Instrument > Handling Diagrams, Traces, and Markers"
in the Help system or in the R&S ZNC Getting Started guide.
Display Settings