Command Reference
R&S
®
ZNC
578
User Manual 1173.9557.02 ─ 13
Example:
CORR:EDEL2:ELEN 0.3
Define an electrical length of 30 cm for channel 1 and port no. 2.
CORR:EDEL2:DIST?; DIEL?
Query the values of the mechanical length and the permittivity at
port 2. The mechanical length is equal to the electrical length divi-
ded by the square root of the permittivity; the latter is set to its
default value. The response is
0.29990704322;1.00062
.
CORR:EDEL2?
Query the value of the delay at port 2. The delay is equal to the
electrical length divided by the speed of light in the vacuum, so the
response is
1.0006922856E-009
.
CORR:LOSS2 2; LOSS2:FREQ 1.5 GHz; OFFS 3 dB
Define the offset loss parameters at port 2.
Manual operation:
See
"Delay / Electrical Length / Mechanical Length"
[SENSe<Ch>:]CORRection:EDELay<PhyPt>[:TIME]
<Delay>
Defines the offset parameter for test port
<PhyPt>
as a delay time.
Suffix:
<Ch>
.
Channel number of the offset-corrected channel
<PhyPt>
Port number of the analyzer
Parameters:
<Delay>
Delay
Range:
-3.40282346038 s to
+3.40282346038 s.
*RST:
0 s
Default unit: s
Example:
See
Manual operation:
See
"Delay / Electrical Length / Mechanical Length"
[SENSe<Ch>:]CORRection:LOSS<PhyPt>
<OffsetLoss>
Defines the offset loss at the reference frequency (
).
Suffix:
<Ch>
.
Channel number of the offset-corrected channel
<PhyPt>
Port number of the analyzer
Parameters:
<OffsetLoss>
Frequency-dependent part of the offset loss
Range:
-200 dB to +200 dB
Increment: 0.001 dB
*RST:
0 dB
Default unit: dB
SCPI Command Reference