Command Reference
R&S
®
ZNC
570
User Manual 1173.9557.02 ─ 13
Example:
CORR:COLL:METH:DEF 'Test',RSHort,1
Select a one-port normalization at port 1 with a short standard as
calibration type.
CORR:COLL:SAVE:SEL:DEF
Calculate a dummy system error correction for the normalization
at port 1. The dummy system error correction provides the reflec-
tion tracking error term 'REFLTRACK'.
CORR:CDAT? 'REFLTRACK',1,0
Query the dummy system error correction term. The response is
a 1 (written as
1,0
for the real and imaginary part) for each sweep
point (no attenuation and no phase shift between the analyzer and
the calibration plane).
INIT:CONT OFF; :INIT; *WAI
Stop the sweep to ensure correct transfer of calibration data.
CORR:CDAT 'REFLTRACK',1,0,<ASCII_data>
Replace the dummy system error correction term with your own
correction data, transferred in ASCII format.
INIT:CONT ON
Restart the sweep in continuous mode.
Usage:
Event
Manual operation:
See
[SENSe<Ch>:]CORRection:COLLect:SAVE:SELected[:DUMMy]
Calculates the system error correction data from the acquired measurement results
(
[SENSe<Ch>:]CORRection:COLLect[:ACQuire]:SELected
), stores it and
applies it to the calibrated channel
<Ch>
. To avoid incompatibilities, older system error
correction data is deleted unless it has been transferred into a cal pool (
<Ch>, '<file_name>'
).
Suffix:
<Ch>
.
Channel number of the calibrated channel
Example:
CORRection:COLLect:METHod:DEFine 'Test',RSHort,
1
Select a one-port normalization at port 1 with a short standard as
calibration type.
CORRection:COLLect:CHANnels:ALL ON
Enable calibration in all channels of the active recall set.
CORRection:COLLect:SELected SHOR,1
Measure a short standard connected to port 1 and store the mea-
surement results of this standard.
CORRection:COLLect:SAVE:SELected
Calculate the system error correction data and apply it to the
channels.
Usage:
Event
Manual operation:
See
SCPI Command Reference