Command Reference
R&S
®
ZNC
446
User Manual 1173.9557.02 ─ 13
Parameters:
<Result>
S | Y | Z
S-parameters, Y-parameters, Z-parameters
Example:
*RST; CALC:PAR:MEAS 'Trc1'", '"Y-S22'
Select the converted admittance Y <-- S22 as measurement
parameter of the default trace.
CALC:TRAN:COMP S
Convert the converted Y-parameter into an S-parameter.
CALCulate<Chn>:TRANsform:IMPedance:RNORmal
<Model>
Selects the theory for the renormalization of port impedances. The selection has an
impact on the conversion formulas for wave quantities and S-parameters.
Suffix:
<Chn>
.
Channel number used to identify the active trace.
Parameters:
<Model>
TWAVes | PWAVes
TWAVes - travelling waves
PWAVes - power waves
*RST:
TWAVes
Example:
See
Manual operation:
See
"Renormalization according to Theory of"
CALCulate<Chn>:TRANsform:TIME:CENTer
<CenterTime>
Defines the center time of the diagram in time domain.
Suffix:
<Chn>
.
Channel number used to identify the active trace.
Parameters:
<CenterTime>
Center time of the diagram in time domain.
Range:
-99.99999999999 s to +99.99999999999 s
Increment: 0.1 ns
*RST:
1.5E-009 s
Default unit: s
Example:
*RST; :CALC:TRAN:TIME:STAT ON
Reset the instrument, activating a frequency sweep, and enable
the time domain transformation for the default trace.
CALC:TRAN:TIME:CENT 0; SPAN 5ns
Set the center time to 0 ns and the time span to 5 ns.
Manual operation:
See
"Time Start / Stop / Center / Span"
Note:
If the x-axis is scaled in distance units (
DISTance
), then the center value is entered in m; the range and default value
changes accordingly.
SCPI Command Reference