Command Reference
R&S
®
ZNC
544
User Manual 1173.9557.02 ─ 13
The different standard types are defined by the following parameters. Port restrictions
are indicated in brackets:
Table 6-6: Standard types and their parameters
<std_type>
Meaning
Parameters in [SENSe<Ch>:]CORRec-
tion:CKIT:<ConnType>:<Standard
-
Type>
MOPen | FOPen
MOPen(P1) |
FOPen(P2) ...
Open (m) or open (f)
Open (m or f with port restriction)...
'<CalkitName>' ... <C3>[, <Port1>]
complete parameter list with capacitance
coefficients, no OPEN | SHORT | MATCh
MSHort | FSHort
Short (m) or short (f)
'<CalkitName>' ... <L3>[, <Port1>]
complete parameter list with inductance
coefficients, no OPEN | SHORT | MATCh
OSHort | MOSHort |
MOSHORT1 | MOSH-
ORT2 | MOSHORT3 |
FOSHort | FOSHORT1
| FOSHORT2 | FOSH-
ORT3
(MOSHort and MOSH-
ORT1 etc. are equiva-
lent)
Offset short (sexless) or offset short (m) or offset short (f)
(for user-defined connector types only)
'<CalkitName>' ... <L3>[, <Port1>]
complete parameter list with inductance
coefficients, no OPEN | SHORT | MATCh
MMTCh | FMTCh
Match (m) or match (f)
'<CalkitName>' ... <MaxFreq>[, <Port1>]
no offset parameters, no load parameters
(polynomial coefficients), no OPEN |
SHORt | MATCh
MSMatch |FSMatch
Sliding match (m) or sliding match (f)
'<CalkitName>' ... <MaxFreq>[, <Port1>]
no offset parameters, no load parameters
(polynomial coefficients), no OPEN |
SHORt | MATCh
MREFlect | FREFlect
Reflect (m) or reflect (f)
'<CalkitName>' ... OPEN | SHORt [,
<Port1>]
no loss, otherwise complete parameter list
MMTHrough |
MFTHrough |
FFTHrough
Through (m - m) or through (m - f) or through (f - f)
'<CalkitName>' ... <Loss>[, <Port1>,
<Port2>]
no load parameters (polynomial coeffi-
cients), no OPEN | SHORt | MATCh
MMLine | MMLINE1 |
MFLine | MFLINE1 |
FFLine | FFLINE1
MMLine(P2P3) ...
Line1 (m - m) or line1 (m - f) or line1 (f - f)
Line1 (m - m, with port restriction)...
'<CalkitName>' ... <Loss>[, <Port1>,
<Port2>]
no load parameters (polynomial coeffi-
cients), no OPEN | SHORt | MATCh
MMLINE2 | MFLINE2 |
FFLINE2
Line2 (m - m) or line2 (m - f) or line2 (f - f), especially for TRL
calibration
'<CalkitName>' ... <Loss>[, <Port1>,
<Port2>]
no load parameters (polynomial coeffi-
cients), no OPEN | SHORt | MATCh
MMLINE3 | MFLINE3 |
FFLINE3
Line3 (m - m) or line3 (m - f) or line3 (f - f), especially for TRL
calibration
'<CalkitName>' ... <Loss>[, <Port1>,
<Port2>]
no load parameters (polynomial coeffi-
cients), no OPEN | SHORt | MATCh
SCPI Command Reference